Comparison of two transcatheter device strategies for occlusion of the patent ductus arteriosus
Authors:Todd M Gudausky  Russel Hirsch  Philip R Khoury  Robert H Beekman
Abstract:Objectives: The present study evaluates two transcatheter closure strategies utilized at a single center and makes recommendations for device selection when occluding the patent ductus arteriosus. Background: A variety of devices are available for transcatheter closure of the patent ductus arteriosus (PDA) but no guidelines exist to guide operator device choice. Methods: A total of 132 patients underwent attempted transcatheter PDA closure utilizing one of two consecutive closure strategies between January 2000 and June 2005. Strategy A (n = 64; January 2000–May 2003) utilized Gianturco coils only. Strategy B (n = 68; June 2003–June 2005) utilized a single Gianturco coil for the PDA with a minimal diameter ≤1 mm (n = 28) or an Amplatzer Duct Occluder (ADO) if the PDA diameter exceeded 1 mm (n = 40). Success was defined as complete occlusion on a follow up echocardiogram. Results: 58 of 64 (90.6%) patients treated utilizing strategy A had successful coil implantation. 68 of 68 (100%) patients treated utilizing strategy B had successful coil/device implantation. At follow up echocardiography, 32 of 44 (72.7%) strategy A patients had complete ductal closure, as compared with 57 of 58 (98.3%) strategy B patients (P < 0.0001). Stepwise logistic regression analysis identified closure strategy as the most powerful predictor of procedural success (OR = 85.9; CI 5.6–9.99). Conclusions: A transcatheter PDA closure strategy consisting of a single Gianturco coil for PDA ≤ 1 mm or an ADO for larger sized PDA (strategy B) achieves superior outcomes compared to the use of coils alone. © 2008 Wiley‐Liss, Inc.
Keywords:pediatric interventions  congenital heart disease  therapy
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号