根尖定位仪在不同指示点测量特点的体外研究
引用本文:席巧玲,李俊,陈晓英,赵星,陈振宇.根尖定位仪在不同指示点测量特点的体外研究[J].北京口腔医学,2013,21(6):338-340.
作者姓名:席巧玲  李俊  陈晓英  赵星  陈振宇
作者单位:518000,深圳 深圳市第三人民医院口腔科;518000,深圳 深圳市第三人民医院口腔科;518000,深圳 深圳市第三人民医院口腔科;518000,深圳 深圳市第三人民医院口腔科;518000,深圳 深圳市第三人民医院口腔科
摘    要:目的研究2种多频根尖定位仪在不同指示点的测量特点。方法截除75颗单根管前磨牙牙冠,GG钻扩大根管冠、中1/3。计算ProPexⅡ和JustyII在显示0和0.5指示点时,定位锉尖与解剖根尖孔间的距离。结果采用Mann-WhitneyU检验分析。结果指示点为0时,ProPexⅡ和JustyⅡ测量值间差异无统计学意义(P〉0.05),中间值分别为0.39mm和0.37mm。指示点为0.5时,ProPexⅡ和JustyⅡ测量差异有统计学意义(P〈0.05),测量中间值分别为0.56mm和0.69mm。同一定位仪在不同参考点下测量值间差异有统计学意义(P〈0.01)。结论本研究条件下,指示点由0变化到0.5时,2种根尖定位仪定位锉尖均逐渐远离解剖根尖孔。指示点0时2种根尖定位仪定位效果相似。指示点0.5时,ProPexⅡ比JustyII定位更接近解剖根尖孔。

关 键 词:根尖定位仪  参考点  解剖根尖孔

Evaluation of two multi-frequency based apex locators with different scales in vitro
Institution:XI Qiao-ling, LI Jun, CHEN Xiao-ying , ZHA O Xing , CHEN Zhen- yu The Stomatology Department of Shenzhen Third People Hospital , Shenzhen 518000, China
Abstract:Objective To evaluate two multi-frequency based apex locators with different scales. Methods Seventy-five extracted mandibular premolar were decoronated at cemento-enamel junction. The coronal and middle part of root canal was flared by GG drills. The distance between the file tips and anatomic foramen was calculated and recorded in 0 and 0. 5 scales displayed by ProPex 11 and Justy II. The results were subjected into Mann-Whitney U analysis. Results The difference between ProPex H and Justy 11 was not significant when the scale point was 0 ( P 〉 0. 05 ), and the average value was 0. 39mm and 0. 37mm. The difference between two locators was significant when the scale point was 0. 5 ( P 〈 0. 05), and the average value was 0. 56mm and 0. 69mm. The measurements obtained under different scales were significantly different by the same apex locator ( P 〈 0. 01 ). Conclusion Under the limit of current study, the file tips determined by two apex locators were located gradually away from the anatomic foramen when the scale point was changed from 0 to 0. 5. The measurements of two locators were not significantly different when the scale point was 0, but file tips determined by ProPex H were located much closer to anatomic foramen compared with that of Justy H when the scale point was 0. 5.
Keywords:Apex locators  Reference  Anatomic foramen