首页 | 本学科首页   官方微博 | 高级检索  
     


Large-volume radiofrequency ablation of ex vivo bovine liver with multiple cooled cluster electrodes
Authors:Haemmerich Dieter  Lee Fred T  Schutt David J  Sampson Lisa A  Webster John G  Fine Jason P  Mahvi David M
Affiliation:Department of Biomedical Engineering, University of Wisconsin, Madison, USA. haemmerich@ieee.org
Abstract:Three methods of creating large thermal lesions with cool-tip cluster electrodes were compared. Three cluster electrodes were arranged 4 cm apart in a triangular array. Eight lesions were created ex vivo in fresh bovine liver (from a butcher) with each method: sequential ablation (three electrodes, 12 minutes each); simultaneous activation of electrodes (12 minutes); and rapid switching of power between electrodes (12 minutes), for which an electronic computer-controlled switch was developed. For sequential, rapid switching, and simultaneous methods, lesion volumes were 137.5 cm(3)+/- 22.2, 116.4 cm(3)+/- 15.2, and 22.3 cm(3)+/- 6.4 (P < .05), respectively, and final temperatures at lesion center were 80 degrees C +/- 5, 97 degrees C +/- 8, and 41 degrees C +/- 3 (P < .001), respectively. Because of electrical interference between electrodes, simultaneous method led to little heating at the center between the electrodes and created small discontinuous lesions. Rapid switching created large round lesions by employing multiple electrodes concurrently, which substantially reduced treatment time and resulted in more effective heating between electrodes.
Keywords:
本文献已被 PubMed 等数据库收录!
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号