首页 | 本学科首页   官方微博 | 高级检索  
     


The use of the knee joint-line balancer to control patella position in revision total knee arthroplasty
Authors:ten Ham Arno M  Wymenga Ate B  Jacobs Wilco C H
Affiliation:Department of Orthopaedics, Sint Maartenskliniek, Hengstdal 3, P.O. Box 9011, Nijmegen 6500 GM, The Netherlands.
Abstract:In revision total knee arthroplasty (TKA), control of the patellar height and the joint line is difficult. Therefore, we developed an adjustable flexion-extension spacer, the Knee Joint-line Balancer (KJB(R)). This device simulates femur component sizes, polyethylene sizes, the joint-line level, and distal femur wedges. The goal of this study is to evaluate the use of an adjustable knee spacer to control patellar height and joint-line during revision total knee surgery. The subjects of the study were the first 10 consecutive patients who had undergone revision of a primary TKA where the KJB was used. A reference group composed of the last 10 patients treated without the use of the KJB was also evaluated. The joint-line position and the patellar height were determined before and after revision TKA. The method described by Figgie et al. was used. The patellar height in the reference group averaged 7.7 mm. Seven of 10 patients had a patella baja, and two of these patients had patellar impingement. One patient needed a proximalisation of the tuberositas. The patellar height in the KJB group averaged 14.6 mm after revision, with only one patient having a patella baja. This new device, adjustable kneespacer for revision TKA ("the KJB"), seems to provide better control of the patellar position in total knee revision.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号