首页 | 本学科首页   官方微博 | 高级检索  
     


Accuracy of Computer Navigation for Acetabular Component Placement in THA
Authors:James A. Ryan  Amir A. Jamali  William L. Bargar
Affiliation:(1) Department of Orthopaedics, Hospital for Special Surgery, New York, NY, USA;(2) Department of Orthopaedic Surgery, UC Davis Medical Center, Sacramento, CA, USA;(3) Joint Surgeons of Sacramento, Sutter General Hospital, 1020 29th Street, Suite 450, Sacramento, CA 95816, USA;
Abstract:The accuracy and precision of any computer-aided surgical device is critical to its utility. We asked the following question: how accurate and precise are the values measured by an imageless computer navigation system as compared with those measured using postoperative CT scans? Twenty-five patients (26 hips) underwent primary THA using an imageless computer navigation system for placement of the acetabular component. Inclination and anteversion were measured in the operative coordinate system as defined by Murray. Accuracy, precision, and bias were computed, and Bland-Altman analysis was used to assess levels of agreement. The accuracy (mean ± standard deviation of the absolute difference between computer-assisted navigation and CT) was 1.8° ± 1.2° for inclination and 2.0° ± 2.0° for anteversion. Precision was 3.4° for inclination and 5.5° for anteversion. Bias was 0.52° for inclination and 0.35° for anteversion. Limits of agreement were 4.26° for inclination and 5.58° for anteversion. An imageless computer navigation system can precisely determine acetabular cup position.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号