Performance of a tonometer for arterial pressure measurement during anesthesia in the elderly |
| |
Authors: | Katsumi Harasawa Masahiro Yamane Yuji Morimoto Osamu Kemmotsu |
| |
Affiliation: | (1) Department of Anesthesiology and Critical Care Medicine, Hokkaido University Graduate School of Medicine, N-15, W-7, Kita-ku, Sapporo 060-8638, Japan |
| |
Abstract: | We investigated performance of an arterial tonometer during anesthesia in the elderly. Thirty patients (17 men and 13 women), ranging from 70 to 86 years old, were included, and agreement between tonometric arterial pressure (TAP) and intraarterial pressure (IAP) was calculated. A total of 6487 paired points was sampled, and values for precision (mean absolute difference ± SD) were 6.8 ± 5.2 for systolic, 6.9 ± 4.6 for mean, and 9.2 ± 5.4mmHg for diastolic blood pressures. Values for bias and limits of agreement [mean difference (TAP – IAP) ± SD, mean difference (TAP – IAP) ± 2 SD, respectively] for systolic, mean, and diastolic pressure were 1.2 ± 8.4 and 1.2 ± 16.8, 5.7 ± 5.9 and 5.7 ± 11.8, and 8.6 ± 6.2 and 8.6 ± 12.4mmHg, respectively. Compared with previous data, aging is likely to affect the performance of an arterial tonometer. |
| |
Keywords: | Tonometry Arterial pressure Aging Atherosclerosis |
本文献已被 PubMed SpringerLink 等数据库收录! |
|