Pentacam HR及Optovue iVue OCT测量圆锥角膜和近视患者角膜厚度的研究
引用本文:李丹,张阳,郭莉莉,王智群,陈可心,李纳. Pentacam HR及Optovue iVue OCT测量圆锥角膜和近视患者角膜厚度的研究[J]. 眼科, 2021, 30(2): 136-141. DOI: 10.13281/j.cnki.issn.1004-4469.2021.02.011
作者姓名:李丹  张阳  郭莉莉  王智群  陈可心  李纳
作者单位:首都医科大学附属北京同仁医院 北京同仁眼科中心 北京市眼科研究所 眼科学与视觉科学北京市重点实验室 100730
摘    要:目的 分析Pentacam HR、Optovue iVue OCT在圆锥角膜和近视患者中角膜厚度测量值的一致性,以期对角膜交联手术和角膜屈光手术患者角膜厚度测量提供指导.设计诊断技术评价.研究对象2018年10月-2020年10月北京同仁医院圆锥角膜和拟行角膜屈光手术的近视患者181例(359眼),其中近视166眼,圆...

关 键 词:圆锥角膜  角膜厚度  Pentacam HR  Optovue iVue OCT
收稿时间:2020-12-23

Study of corneal thickness measurements with Pentacam HR and Optovue iVue OCT in keratoconus and myopic patients
Li Dan,Zhang Yang,Guo Lili,Wang Zhiqun,Chen Kexin,Li Na. Study of corneal thickness measurements with Pentacam HR and Optovue iVue OCT in keratoconus and myopic patients[J]. Ophthalmology in China, 2021, 30(2): 136-141. DOI: 10.13281/j.cnki.issn.1004-4469.2021.02.011
Authors:Li Dan  Zhang Yang  Guo Lili  Wang Zhiqun  Chen Kexin  Li Na
Affiliation:Beijing Tongren Eye Center, Beijing Institute of Ophthalmology, Beijing Tongren Hospital, Capital Medical University, Beijing Key Laboratory of Ophthalmology and Visual Science, Beijing 100730, China
Abstract:Objective To analyze the consistency of Pentacam HR and Optovue iVue OCT in the measurement of corneal thickness in patients with keratoconus and myopia,and to provide guides for the measurement of corneal thickness in keratoconus and myopic patients.Design Diagnostic technique evaluation.Participants 181 cases(359 eyes)of keratoconus and myopia patients scheduled for refractive surgery from October 2018 to October 2020 in Beijing Tongren Hospital.166 eyes were in the myopia group,and 193 eyes were in the keratoconus group.Methods Pentacam HR and Optovue iVue OCT were used to measure the thickness and related parameters of different cornea areas.A one-way analysis of variance or Kruskal-Wallis H test was used to analyze the differences.the Bland-Altman consistency test was used to analyze the consistency.The Spearman correlation coefficient was used to analyze the correlation between the differences and corneal surface morphological parameters.Main Outcome Measures Corneal thickness in central corneal thickness,the thinnest corneal thickness,maximum keratommetry(Kmax),index of surface variance(ISV),index of vertical asymmetry(IVA),index of height asymmetry(IHA),index of height decentration(IHD).Results In the myopia group and keratoconus group,the thinnest point and central thickness of the cornea measured by Optovue iVue OCT was about 20μm lower than Pentacam HR on average,and the difference was statistically significant(all P<0.05).For the central and thinnest thickness in myopia group and keratoconus group,the consistency of the measurement results of the two measuring instruments are quite different.The difference both the central area and the thinnest part in myopia group was about 35μm.The measured difference spans of the central area and the thinnest part in keratoconus group were 60μm and 84.56μm,respectively.Both myopia group and keratoconus group,the corneal thickness difference in the central area and the thinnest point had no obvious linear correlation with KMax,ISV,IVA,IHA,IHD.The central corneal thickness in myopia group:r=0.025,-0.15,-0.165,-0.104,-0.15;PIVA=0.034.However,the scatter plot showed no linear correlation,and the other P values all were>0.05.Conclusions The corneal thickness measured with Optovue iVue OCT is about 20μm lower than with Pentacam HR.It is recommended to refer to the corneal thickness measured with Optovue iVue OCT for patients before refractive surgery and keratoconus cross-linking surgery.
Keywords:keratoconus,corneal thickness,Pentacam HR  Optovue iVue OCT