Biocompatibility and pathways of initial complement pathway activation with Phisio- and PMEA-coated cardiopulmonary bypass circuits during open-heart surgery |
| |
Authors: | Thiara A S Mollnes T E Videm V Andersen V Y Svennevig K Kolset S O Fiane A E |
| |
Affiliation: | Department of Thoracic and Cardiovascular Surgery, Oslo University Hospital, Oslo, Norway. amrit.thiara@rikshospitalet.no |
| |
Abstract: | A randomized open-heart surgery study comprising 30 patients was undertaken to compare the biocompatibility of Phisio-(phosphorylcholine) and PMEA-(poly-2-methoxyethyl acrylate) coated cardiopulmonary bypass (CPB) circuits and to assess the initial complement pathway activation during open-heart surgery. Blood samples were obtained at five time points, from the start of surgery to 24 hours postoperatively. The following analyses were performed: haemoglobin, lactate dehydrogenase, leukocyte and platelet counts, myeloperoxidase and neutrophil-activating peptide-2, thrombin-anti-thrombin complexes, syndecan-1 and the complement activation products C1rs-C1-inhibitor complexes, C4bc, C3bc, C3bBbP and the terminal complement complex (TCC). No significant inter-group difference was found in any parameters, except for the concentration of TCC which was moderately lower in the PMEA group at termination of CPB. Complement activation during open-heart surgery was mainly mediated through the alternative pathway. In conclusion, PMEA- and Phisio-coated circuits displayed similar biocompatibility with respect to inflammatory and haemostatic responses during and after open-heart surgery. |
| |
Keywords: | |