首页 | 本学科首页   官方微博 | 高级检索  
检索        


Measurement and correction of transverse chromatic offsets for multi-wavelength retinal microscopy in the living eye
Authors:Wolf M Harmening  Pavan Tiruveedhula  Austin Roorda  Lawrence C Sincich
Institution:1University of California, Berkeley, School of Optometry, Berkeley, CA 94720, USA;2University of Alabama at Birmingham, Department of Vision Sciences, Birmingham, AL 35294, USA
Abstract:A special challenge arises when pursuing multi-wavelength imaging of retinal tissue in vivo, because the eye’s optics must be used as the main focusing elements, and they introduce significant chromatic dispersion. Here we present an image-based method to measure and correct for the eye’s transverse chromatic aberrations rapidly, non-invasively, and with high precision. We validate the technique against hyperacute psychophysical performance and the standard chromatic human eye model. In vivo correction of chromatic dispersion will enable confocal multi-wavelength images of the living retina to be aligned, and allow targeted chromatic stimulation of the photoreceptor mosaic to be performed accurately with sub-cellular resolution.OCIS codes: (110.1080) Active or adaptive optics, (130.2035) Dispersion compensation devices, (170.5810) Scanning microscopy, (330.5510) Psychophysics, (330.7327) Visual optics, ophthalmic instrumentation