肝素修饰人工晶体植入在抗青光眼术后白内障超声乳化术中的应用
引用本文:王金华,刘剑萍,张虹,王宁利,王晓贞.肝素修饰人工晶体植入在抗青光眼术后白内障超声乳化术中的应用[J].中国实用眼科杂志,2007,25(1):57-60.
作者姓名:王金华  刘剑萍  张虹  王宁利  王晓贞
作者单位:1. 434020,荆州,华中科技大学同济医学院附属荆州医院眼科
2. 华中科技大学同济医学院附属同济医院眼科
3. 北京同仁眼科中心
摘    要:目的观察肝素表面修饰折叠型人工晶体植入对抗青光眼术后白内障超声乳化吸除术后前房炎症反应的影响。方法对60例(88眼)抗青光眼术后白内障患者施行超声乳化吸除术,所有病例随机植入肝素表面修饰折叠型人工晶体或蓝光滤过折叠型人工晶体。应用激光闪光细胞检测仪(Laserflarecellmeter,LFCM)测量并比较术前及术后1天、7天、30天及90天房水的平均闪辉值,以评价前房炎症反应。并观察术前后的视力、眼压及滤过泡和角膜内皮计数情况。结果(1)术前,二组患者房水闪辉值没有显著性差异,植入肝素表面修饰折叠型人工晶体者,手术后7天内的房水闪辉值均明显低于蓝光滤过折叠型人工晶体植入组,而术后30、90天时两种人工晶体植入者之间差异无显著性。两组术后视力均明显提高,眼压控制稳定,角膜内皮功能正常。结论抗青光眼术后白内障手术后眼血-房水屏障功能受损,植入肝素表面修饰折叠型人工晶体能显著减轻白内障术后短期内的炎症反应,增加了手术安全性。

关 键 词:血-房水屏障  肝素表面修饰  超声乳化  抗青光眼术  白内障
修稿时间:2006-04

The clinical study of implantation of heparin-surface-modified foldable intraocular lens in cataract after glaucoma surgery
WANG Jinhua, LIU Jianping, ZHANG Hong, et al.The clinical study of implantation of heparin-surface-modified foldable intraocular lens in cataract after glaucoma surgery[J].Chinese Journal of Practical Ophthalmology,2007,25(1):57-60.
Authors:WANG Jinhua  LIU Jianping  ZHANG Hong  
Institution:Department of ophthalmology, Jing zhou Hospital, Tongji Medioal College, Huazhong University of Science and Technology, Jing zhou 134020, China
Abstract:Objective To observe the inflammation in anterior chamber after phacoemulsificationwith implantation 2 types of foldable intraocular lens(IOL) in cataract patients after glaucoma surgery.Meth-ods The phacoemulsification was performed in 88 eyes of 60 cataract patients after glaucoma surgery andheparin-surface-modified foldable IOL and sigle-piece acrylic UV block foldable IOL were implantated randomly.The aqueous humor flare value was examined preoperatively and on day 1,7,30 and 90 after surgery with a laserflare cell meter.Visual acutity(VA) ,filter bleb,introcular pressure(IOP) and corneal endothelium was examined.Results Before surgery,there was no significiant difference between 2 groups.Flare was significiantly lowerin patients with implantation of heparin-surface-modified IOL on day 7 postoperatively than patients with im-plantation of single-piece acrylic UV block foldable IOL,there was no significiant difference on day 30,90postoperatively.VA,filter bleb and IOP was well in all patients postoperatively,the counts of corneal endothe-lium was in normal range.Conclusions The blood-aqueous barrier was damaged in patients after glaucomasurgery.The inflammation in anterior chamber can be reduced with implantation of heparin-surface-modifiedIOL.
Keywords:blood-aqueous barrier  heparin-surface-modified IOL  phacoemulsification  glau-coma surgery  cataract