A pulsatile pump for cardiopulmonary bypass and its clinical use
Authors:Yoshitaka Sasaki MD  Takahiro Kawai  Katsuhiko Nishiyama  Yuichiro Murayama  Shogo Toda  Tukio Wada  Kazuhiro Kitaura  Shinichi Sato  Masaharu Kadowaki  Yutaka Kanki  Takafumi Hashimoto  Shuji Shirakata  Koichi Oga  Takahiro Oka
Affiliation:(1) The Division of Surgery, Children’s Research Hospital, Kyoto Prefectural University of Medicine, Kyoto, Japan;(2) The Department of Surgery, Kyoto Prefectural University of Medicine, 465 Kajii-cho, Kawaramachi-Hirokoji, Kamikyo-ku, 602 Kyoto, Japan
Abstract:A pulsatile pump driven by a coil spring, which was designed and constructed by us, is described in this report. It consists of two main parts, a disposable blood chamber and a driving section. The blood chamber has two leaflet valves and a piston, which is covered with two bellofram rolling diaphragms and moves into the housing to draw in and eject the blood. The driving section consists of three cams, an electric motor and a coil spring. The ejection force is wholly produced by the compressed coil spring and is transmitted to the piston in the blood chamber by a rod. This pump allows the ejection pressure, the beat rates, and the stroke volume all to be changed independently. The performance of the pump was tested by using a circulation model where the beat rate was adjusted from 30 to 250 bpm. The output subsequently increased from 0.81/min to 5.71/min and the stroke volume, from 20.4 ml to 36.7 ml. This new pump has been used for clinical cardiopulmonary bypasses in 24 patients of open heart surgery and the pressure traces during perfusion resembled those of the patients’ own hearts.
Keywords:cardiopulmonary bypass  pulsatile pump  bellofram rolling diaphragm
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号