4种便携式血糖仪与生化分析仪的一致性和抗干扰能力的研究
引用本文:吴小利,吴杨,张云丽,娜仁高娃,薄帅,巩丽颖. 4种便携式血糖仪与生化分析仪的一致性和抗干扰能力的研究[J]. 国际检验医学杂志, 2022, 43(5): 534-538,544. DOI: 10.3969/j.issn.1673-4130.2022.05.005
作者姓名:吴小利  吴杨  张云丽  娜仁高娃  薄帅  巩丽颖
作者单位:辽宁省人民医院(中国医科大学人民医院)检验医学科,辽宁沈阳110016,辽宁省人民医院(中国医科大学人民医院)重症医学科,辽宁沈阳110016
基金项目:辽宁省自然科学基金指导计划项目;辽宁省科学技术计划项目
摘    要:目的 通过对不同便携式血糖仪与贝克曼库尔特A U 5800全自动生化分析仪(A U 5800)检测血糖结果的一致性比对和抗干扰能力的研究,为临床选择适合的血糖仪提供参考.方法 共选取4种不同检测原理的2台进口(A和B)和2台国产(C和D)血糖仪,与A U 5800同时检测不同浓度的血糖并进行比较,分析其一致性.分别检测...

关 键 词:便携式血糖仪  红细胞比容  一致性  抗干扰能力

Comparison of the consistency among four portable blood glucose meters and biochemical analyzer and study of their anti-interference ability
WU Xiaoli,WU Yang,ZHANG Yunli,NA Rengaowa,BO Shuai,GONG Liying. Comparison of the consistency among four portable blood glucose meters and biochemical analyzer and study of their anti-interference ability[J]. International Journal of Laboratory Medicine, 2022, 43(5): 534-538,544. DOI: 10.3969/j.issn.1673-4130.2022.05.005
Authors:WU Xiaoli  WU Yang  ZHANG Yunli  NA Rengaowa  BO Shuai  GONG Liying
Affiliation:(Department of Laboratory Medicine,the People′s Hospital of Liaoning Province(the People′s Hospital of China Medical University),Shenyang,Liaoning 110016,China;Intensive Care Unit,the People′s Hospital of Liaoning Province(the People′s Hospital of China Medical University),Shenyang,Liaoning 110016,China)
Abstract:Objective To study on the consistency between different portable blood glucose meter and beckman-coulter AU5800 automatic biochemical analyzer and their anti-interference ability and to provide reference for clinical selection of suitable blood glucose meter.Methods Two imported(A and B)and two home-made(C and D)blood glucose meters with four different detection principles were selected to detect different concentrations of blood glucose simultaneously with AU5800.Three samples of hematocrit(HCT)levels of low,medium,and high blood glucose were measured with blood glucose meters.Vitamin C(VC),galactose,maltose and bilirubin with different concentrations were selected to observe whether the anti-interference ability of each glucose meter met the requirements.Results The two imported blood glucose meters had a good consistency with AU5800 and met the requirements of ISO 15197:2013,but the consistency of the two home-made blood glucose meters could not meet the requirements of the standards.The coincidence rates of home-made C and D were 88.00%and 92.00%,respectively.Compared with HCT 45.00%of sample,the bias of home-made C in HCT 25.00%and 65.00%were 20.22%,-31.93%,31.03%and-24.39%in the middle and high blood glucose samples,the differences were statistically significant(P<0.05);the bias of home-made D in HCT 25.00%and 65.00%were 47.50%,-28.79%,46.52%and-30.05%,the difference were statistically significant(P<0.05).In the anti-interference study,when VC were 0.5 g/L and 1.0 g/L,the bias of home-made C were 0.60 mmol/L and 1.31 mmol/L in low glucose,the differences were statistically significant(P<0.05),while the results of other glucose meters showed no statistical difference.Conclusion The results of imported A and B glucose meters were in good agreement.The results of home-made C and D were influenced by HCT,and positive deviation was found in medium HCT,negative deviation was found in high HCT.VC interferes with the measurement of home-made C.Therefore,the glucose meter should be selected according to the needs to avoid the influence of endogenous and exogenous distractors.
Keywords:portable glucose meter  hematocrit  consistency  anti-interference