Evaluation of semiautomated internal carotid artery stenosis quantification from 3-dimensional contrast-enhanced magnetic resonance angiograms
Authors:van Bemmel Cornelis M  Elgersma Otto E H  Vonken Evert-Jan P A  Fiorelli Marco  van Leeuwen Maarten S  Niessen Wiro J
Affiliation:University Medical Center, Image Sciences Institute, Utrecht, Netherlands.
Abstract:RATIONALE AND OBJECTIVES: The performance of a semiautomatic technique for internal carotid artery (ICA) stenosis quantification of the internal carotid artery in contrast-enhanced magnetic resonance angiography was evaluated. MATERIALS AND METHODS: The degree of stenosis of 52 ICAs was quantified by measuring the cross-sectional area along the center lumen line. This was performed both by 3 independent observers and the semiautomated method. The degree of stenosis was defined as the amount of cross-sectional lumen reduction. RESULTS: Agreement between the method and observers was good (weighted-kappa, kappaW = 0.89). Reproducibility of measurements of the semiautomated technique was better (kappaW = 0.97) than that of the observers (kappaW = 0.76), and the evaluated technique was considerably less time-consuming. CONCLUSIONS: Because the user interaction is limited, this technique can be used to replace an expert observer in 3-dimensional stenosis quantification of the ICA at CE-MRA in clinical practice.