电感耦合等离子体质谱仪与国标法测水碘的方法比较 |
| |
引用本文: | 招莉,孙晓帆.电感耦合等离子体质谱仪与国标法测水碘的方法比较[J].疾病监测与控制杂志,2014(7):412-413. |
| |
作者姓名: | 招莉 孙晓帆 |
| |
作者单位: | 韶关市疾病预防控制中心,广东韶关512028 |
| |
摘 要: | 目的比较电感耦合等离子体质谱法与国标法的水碘检测结果。方法从方法的灵敏度,精密度和准确度等指标进行比较。结果电感耦合等离子体质谱法测定水碘检出限0.2μg/L,加标回收率95.0%~102.0%,相对标准偏差2.0%,两种方法对实际样品的分析结果没有显著统计学意义。结论电感耦合等离子体质谱法操作简单、快速、准确,是一种检测水碘的较好方法。
|
关 键 词: | 电感耦合等离子体质谱 水 碘 |
Comparative mothod to measure the water iodine by ICP-MS and national standard method |
| |
Authors: | ZHAO Li SUN Xiao-fan |
| |
Institution: | (Shaoguan Center for Disease Control and Prevention, Shaoguan 512028, China) |
| |
Abstract: | Objective Comparision of the results of water iodine by ICP-MS and national standard method. Methods Compared the sensitivity of the method,accuracy and precision. Results Detection limit of the method was 0.2;g/L, the recovery is 95.0%;102.0%, the relative standard deviation is less than 2%, two methods for analysis of real samples were not significant difference. Conclusion ICP-MS is a simple,rapid,accurate, is a good method to detect the water iodine. |
| |
Keywords: | ICP-MS Water Iodine |
本文献已被 CNKI 维普 等数据库收录! |
|