UF-50全自动尿沉渣分析仪检测尿中管型结果的分析
引用本文:彭瑛,万红,王开正,叶婷,梁军.UF-50全自动尿沉渣分析仪检测尿中管型结果的分析[J].泸州医学院学报,2009,32(5):527-529.
作者姓名:彭瑛  万红  王开正  叶婷  梁军
作者单位:泸州医学院附属医院检验科,四川泸州,646000
摘    要:目的:探讨UF-50全自动尿沉渣分析仪检测尿中管型的准确性及其影响因素。方法:随机收取住院患者的晨尿标本600份,同时进行UF-50尿液沉渣自动化分析(简称UF-50)和显微镜(镜检)检测。将两种方法检测的管型结果进行比较分析。结果:600份标本中,UF-50检测阳性的有200例,而镜检阳性的有39例,假阳性率为29.1%;UF-50检测阴性尿标本400例,而镜检阳性的有7例,假阴性率为15.2%。结论:使用UF-50检测尿液管型存在着较高的假阳性率和假阴性率。假阳性结果主要由上皮细胞、结晶、黏液丝、脓细胞等有形成分引起。因此对UF-50检测管型阳性的尿标本应用镜检进行确证,而对管型阴性但尿蛋白阳性的尿标本也应该用镜检进行排除,以免漏诊。

关 键 词:尿液  尿沉渣分析仪  管型

RESULTS ANALYSIS OF URINARY CAST DETECTED BY UF-50 COMPLETE AUTOMATIC URINARY SEDIMENT ANALYZER
Institution:Peng Ying, et al (Department of Laboratory Medicine, Affiliated Hospital of Luzhou Medical College)
Abstract:Objective:To study the accuracy and influencing factors of UF-50 complete automatic urinary sediment analyzer for detecting the urinary casts. Methods: UF-50 complete automatic urinary sediment analyzer and microscope were used to examine the casts in 600 morning urine samples randomly collected from in-patients .The two results were analyzed and compared. Results:In 200 positive cases examined by UF-50, only39 were confirmed by microscopic examination, the false-positive rate was 29.1%;in 400 negative cases examined by UF-50, 7 positive cases were detected by microscopic examination, the false-negative rate was 15.2%. Epithelium, crystal, slime filament, pus cells were always found in these false positive cases by microscopic examination. Conclusion:There are higher false-positive rate and false-negative rate in the examination of urinary casts by UF-50 complete automatic urinary sediment analyzer. Therefore,these casts positive urine specimen examined by UF-50 Meanwhile,the casts negative urine specimen also should be checked by microscope when the urine protein is positiye.should be checked by mieroscop .
Keywords:Urine  Urine sediment analyzer  Casts