TLD-LiF热释光剂量计在冠状动脉血管内再狭窄剂量学的应用研究
引用本文:王建华,徐志勇,张良安,戴光复,苑淑渝. TLD-LiF热释光剂量计在冠状动脉血管内再狭窄剂量学的应用研究[J]. 中国辐射卫生, 2004, 13(4): 241-243. DOI: 10.13491/j.cnki.issn.1004-714x.2004.04.001
作者姓名:王建华  徐志勇  张良安  戴光复  苑淑渝
作者单位:1. 常州肿瘤医院, 江苏常州 213000;2. 复旦大学附属肿瘤医院;3. 中国医学科学院放射医学研究所
基金项目:国家自然科学基金资助项目批准号 :30 1 70 2 33
摘    要:目的 研究粉末,3 mm×3 mm×0.4 mm TLD方片,3 mm×3 mm×0.2 mm TLD方片和1 mm×1 mm×1 mm TLD方片4种TLD-LiF剂量计在近距离放射治疗冠状动脉再狭窄时的应用特性。方法 用上述4种剂量计测量32P液体球囊(直径为3 mm,长度为20 mm)和103Pd放射性支架(直径为20 mm,长度为13 mm)的表面剂量和径向剂量,并将测量结果与蒙特卡罗代码(MCNP4B)的计算结果进行比较。结果 粉末和方片剂量计的测量结果有一定的差异,经过方片粉末修正因子修正后,3 mm×3 mm×0.2 mm TLD方片的结果与粉末一致。结论 3 mm×3 mm×0.2 mmTLD方片可以测量冠状动脉再狭窄近距离放射治疗的径向剂量和表面剂量分布。

关 键 词:TLD热释光剂量计  放射性支架  再狭窄  
文章编号:1004-714X(2004)04-0241-02
收稿时间:2003-12-21
修稿时间:2003-12-21

Application Study for LiF Thermoluminescence Dosimeter in Intracoronary Brachytherapy Restenosis of Coronary Artery
WANG Jian-hua,XU Zhi-yong,ZHANG Liang-an,et al.. Application Study for LiF Thermoluminescence Dosimeter in Intracoronary Brachytherapy Restenosis of Coronary Artery[J]. Chinese Journal of Radiological Health, 2004, 13(4): 241-243. DOI: 10.13491/j.cnki.issn.1004-714x.2004.04.001
Authors:WANG Jian-hua  XU Zhi-yong  ZHANG Liang-an  et al.
Affiliation:Changzhou Tumor Hospital, Jiangsu, China
Abstract:Objective To investigate application characteristic for powder, 3 mm×3 mm×0.4 mm, 3 mm×3 mm×0.2 mm and 1 mm×1 mm×1 mm TLD-LiF chips dosimeters intracoronary brachytherapy restenosis of coronary artery.Methods Dose distribution of surface and radial for a 103Pd radioactive stent and a 32P radioactive liquid-filled balloon were measured by above-mentioned four kinds of LiF thermoluminescence dosimeters (TLD).The results were compared to those calculated by MCNP4B code.Results Difference of dose measurement was shown between powders and various chip TLDs.After the results from 3mm×3mm×0.2 mm chip TLDs were revised by chip-powder adjustment factor, results from this chip and powders was found to show good agreement within the error of the calculations (a few per cent).Conclusion Dose distribution of surface and radial for intracoronary brachytherapy restenosis of coronary artery can be measured by 3 mm×3 mm×0.2 mm TLD chips.
Keywords:Thermoluminescence Dosimeters  Radioactivity Stent  Restenosis