Intraoperative patient warming using radiant warming or forced-air warming during long operations
Authors:Lee L  Leslie K  Kayak E  Myles P S
Affiliation:Department of Anaesthesia and Pain Management, The Royal Melbourne and The Alfred Hospital, Melbourne, Victoria.
Abstract:Forced-air warming is the most commonly used and effective method of active warming. A new radiant warming device (Suntouch, Fisher and Paykel) may provide an alternative when the skin surface available for warming is limited. We conducted a randomized controlled trial to compare the efficacy of the Suntouch radiant warmer and forced-air warming. With ethics committee approval 60 surgical patients having procedures anticipated to be more than two hours in duration were recruited. Patients were randomized to either radiant warming or forced-air warming. All intravenous fluids were warmed but prewarming was not used. The final intraoperative core temperatures (degrees C) for the radiant warming and forced-air warming groups were 36.0 +/- 0.5 and 36.4 +/- 0.6 (P=0.002) respectively. No other patient variables were significantly different. The Suntouch is not as effective as the forced air warming for intraoperative warming during long surgical procedures. The device may be useful when forced-air warming is not possible.
Keywords:
本文献已被 PubMed 等数据库收录!