首页 | 本学科首页   官方微博 | 高级检索  
     


Image quality of supine chest radiographs: intra-individual comparison of computed radiography and low-dose linear-slit digital radiography
Authors:Zsolt Szucs-Farkas  Peter Vock
Affiliation:(1) Department of Diagnostic, Interventional and Paediatric Radiology, Inselspital, Berne University Hospital and University of Berne, Freiburgstrasse 4, Berne, 3010, Switzerland
Abstract:The purpose of this retrospective study was to intra-individually compare the image quality of computed radiography (CR) and low-dose linear-slit digital radiography (LSDR) for supine chest radiographs. A total of 90 patients (28 female, 62 male; mean age, 55.1 years) imaged with CR and LSDR within a mean time interval of 2.8 days ± 3.0 were included in this study. Two independent readers evaluated the image quality of CR and LSDR based on modified European Guidelines for Quality Criteria for chest X-ray. The Wilcoxon test was used to analyse differences between the techniques. The overall image quality of LSDR was significantly better than the quality of CR (9.75 vs 8.16 of a maximum score of 10; p < 0.001). LSDR performed significantly better than CR for delineation of anatomical structures in the mediastinum and the retrocardiac lung (p < 0.001). CR was superior to LSDR for visually sharp delineation of the lung vessels and the thin linear structures in the lungs. We conclude that LSDR yields better image quality and may be more suitable for excluding significant pathological features of the chest in areas with high attenuation compared with CR.
Keywords:Digital radiography  Computed radiography  Image quality  Chest X-ray
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号