首页 | 本学科首页   官方微博 | 高级检索  
     


A method for bone-cement interface thermometry. An in vitro comparison between low temperature curing cement Palavit, and Surgical Simplex P.
Authors:S Harving  K S?balle  C Bünger
Affiliation:Biomechanics Laboratory, Orthopedic Hospital, Arhus, Denmark.
Abstract:Previous temperature measurements at the bone-cement intreface have all shown large variations. To evaluate more precisely the temperature profiles during cement curing, a new experimental model was developed. Eight thermocouple electrodes at the bone-cement interface for each test specimen were used for continuous temperature recordings. Temperature profiles of Palavit was compared with those of Surgical Simplex P in an in vitro model using isolated pig femurs. Defects of 12 x 17 mm in the femoral metaphysis were filled with cement. In six tests with each type of cement, Palavit peaked at a temperature of 50 +/- 0.5 degrees C, whereas Surgical Simplex P peaked at 60 +/- 0.7 degrees C. Core temperatures reached peak values of 70 +/- 0.8 degrees C and 95 +/- 2.2 degrees C for Palavit and Surgical Simplex P, respectively.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号