Goldmann眼压计与非接触眼压计测量青光眼患者眼压的对比分析
引用本文:李红锋,欧阳君,屈晓勇.Goldmann眼压计与非接触眼压计测量青光眼患者眼压的对比分析[J].国际眼科杂志,2015,15(1):144-145.
作者姓名:李红锋  欧阳君  屈晓勇
作者单位:1. 332000,中国江西省九江市妇幼保健院
2. 332000,中国江西省九江市第一人民医院
摘    要:目的::对比分析 Goldmann 压平式眼压计( Goldmann applanation tonometer, GAT)和非接触眼压计( non-contact tonometer,NCT)测量青光眼患者眼压的结果,寻找其规律性。方法:分别使用GAT和NCT测量108例206眼青光眼患者的眼压。结果:青光眼患者108例206眼的平均眼压:GAT测量为29.77±10.27mmHg,NCT测量为24.59±8.58mmHg,两者比较存在显著性差异(P<0.01),而且眼压越高,差异性越显著。结论:NCT 测量值较 GAT 测量值低,眼压越高,差值越大。建议使用GAT检查青光眼患者的眼压,以免漏诊和贻误治疗。

关 键 词:非接触式眼压计  Goldmann压平眼压计  青光眼
收稿时间:2014/5/17 0:00:00
修稿时间:2014/12/19 0:00:00

Comparison of two methods of tonometry in glaucoma patients: Goldmann applanation tonometer and non-contact tonometer
Hong-Feng Li,Jun Ouyang and Xiao-Yong Qu.Comparison of two methods of tonometry in glaucoma patients: Goldmann applanation tonometer and non-contact tonometer[J].International Journal of Ophthalmology,2015,15(1):144-145.
Authors:Hong-Feng Li  Jun Ouyang and Xiao-Yong Qu
Institution:The Maternal and Child Health Hospital of Jiujiang, Jiujiang 332000, Jiangxi Province, China;The First People's Hospital of Jiujiang, Jiujiang 332000, Jiangxi Province, China;The First People's Hospital of Jiujiang, Jiujiang 332000, Jiangxi Province, China
Abstract:AIM: To compare the two methods for the measurement of glaucoma patients' intraocular pressure(IOP)between Goldmann applanation tonometer(GAT)and non-contact tonometer(NCT)and find the laws of the two methods.

METHODS: The IOP of 108 glaucoma patients(206 eyes)were measured by GAT and NCT respectively.

RESULTS: In 108 glaucoma patients, the average IOP of 206 eyes was 29.77±10.27mmHg by GAT and 24.59±8.58mmHg by NCT. There was significant difference between GAT and NCT(P<0.01). The higher of IOP, the difference between GAT and NCT was greater.

CONCLUSION: The measurement results with NCT were lower than that of GAT. The higher of IOP, the difference between GAT and NCT was greater. It's better to measure IOP by GAT for the glaucoma patients, in order to avoid the misdiagnosis and mistreatment of glaucoma.

Keywords:non-contact tonometer  Goldmann applanation tonometer  glaucoma