Measurement of current spread from microelectrodes when stimulating within the nervous system |
| |
Authors: | E. V. Bagshaw M. H. Evans |
| |
Affiliation: | (1) A.R.C. Institute of Animal Physiology, CB2 4AT Babraham, Cambridge, England |
| |
Abstract: | Summary Tungsten stimulating microelectrodes have been tested in monopolar, bipolar and concentric configurations for the extent to which unwanted current spread occurred. Current spread from monopolar electrodes in close conformity with the predictions of the inverse square law, both in vitro and in vivo. The bipolar and concentric configurations, tested only in vitro, had current-spread characteristics which did not follow the inverse square law so closely. The bipolar configuration gave little reduction in spread, compared with a monopolar electrode, but the concentric configuration did reduce the extent of stimulus spread. The extent of spread depends greatly upon a number of experimental variables, including the dimensions of the microelectrode tip. For studies requiring precise localization, with low stimulus currents, it is advisable to determine the inverse square law constant for each experimental situation. For more generalized stimulation purposes within the mammalian C.N.S., employing monopolar microelectrodes of moderate tip size, brief pulses of 10 A can be expected to stimulate myelinated axons within about 0.15 mm of the tip, and 100 A will stimulate within a radius of about 0.5 mm.The work was all done while E.V.B. was on Industrial Training leave from the North-East Polytechnic, Romford Road, London E. 15 |
| |
Keywords: | Current spread Microelectrodes Stimulus spread |
本文献已被 SpringerLink 等数据库收录! |
|