首页 | 本学科首页   官方微博 | 高级检索  
检索        


A parametric analysis of fixation post shape in tibial knee prostheses
Authors:Au Anthony G  Liggins Adrian B  Raso V James  Amirfazli A
Institution:Department of Mechanical Engineering, University of Alberta, 4-9 Mechanical Engineering Building, Edmonton, Alta., Canada T6G 2G8.
Abstract:A primary concern of total knee replacement (TKR) is aseptic loosening of the tibial component, which may be caused by shielding of mechanical stresses in the bone and may require subsequent revision surgery. A three-dimensional (3D) finite element (FE) model has been developed to study bone and interface stresses for four different tibial prosthesis designs. The model described here incorporates orthotropic and heterogeneous bone properties with physiologically representative loading conditions. Results from this model indicate that stress distribution is affected by the incorporation of anisotropy and spatial variation of bone properties. All bone properties were mapped from published data to characterize their anisotropy and heterogeneity. Physiological loading was incorporated by mapping experimentally determined contact patterns. Convergence testing was performed to ensure model accuracy. In terms of interface forces, a tapered post decreased post shear while slightly increasing post compression compared to a cylindrical post; a post of elliptical cross-section increased post shear and decreased post compression. In terms of cancellous bone stress, tapered and elliptical posts both relieved compression compared to a cylindrical post, while a tapered post also produced increased peripheral stress. The inclusion of medial and lateral pegs in addition to a central fixation post caused localized stress shielding in the periphery of the pegs. In general, all implant models caused a reduction of cancellous bone stress plus high compression beneath the central fixation posts.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号