上海市16台医用电子加速器输出量的质量控制
引用本文:赵静芳,易艳玲,张辉,孔燕,高轶名,卓维海. 上海市16台医用电子加速器输出量的质量控制[J]. 环境与职业医学, 2009, 26(6): 553-555
作者姓名:赵静芳  易艳玲  张辉  孔燕  高轶名  卓维海
作者单位:赵静芳 (复旦大学放射医学研究所,上海,200032) ; 易艳玲 (复旦大学放射医学研究所,上海,200032) ; 张辉 (复旦大学放射医学研究所,上海,200032) ; 孔燕 (复旦大学放射医学研究所,上海,200032) ; 高轶名 (复旦大学放射医学研究所,上海,200032) ; 卓维海 (复旦大学放射医学研究所,上海,200032) ;
基金项目:上海市放射卫生与放射病防治重点学科建设项目,上海市卫生局科研项目
摘    要:[目的]了解上海市放射治疗医用电子加速器输出量的质量控制水平。[方法]将核查用的剂量计置于30cm×30cm×30cm水模体的水面下5cm中心处,在源点距水面100cm、照射野10cm×10cm的条件下,给予约2Gy吸收剂量的照射;计算所给予剂量与剂量计测量值的相对偏差。相对偏差≤土5%为合格。[结果]所调查的16台医用电子加速器,其预设剂量与实测剂量的相对偏差范围为-6.8%~8.1%,总合格率为87.5%。有2台机器的相对偏差超过±5%。[结论]上海市放射治疗医用电子加速器输出量合格率的总体情况良好,但仍需要加强对放射治疗技术人员的业务素质培训和监督核查。

关 键 词:放射治疗剂量  质量控制  玻璃剂量计

Quality Control of Radiation Output of 16 Medical Accelerators in Shanghai
ZHAO Jing-fang,YI Yan-ling,ZHANG Hui,KONG Yan,GAO Yi-ming,ZHUO Wei-hai. Quality Control of Radiation Output of 16 Medical Accelerators in Shanghai[J]. Journal of Environmental & Occupational Medicine, 2009, 26(6): 553-555
Authors:ZHAO Jing-fang  YI Yan-ling  ZHANG Hui  KONG Yan  GAO Yi-ming  ZHUO Wei-hai
Affiliation:( Institute of Radiation Medicine, Fudan University, Shanghai 200032, China )
Abstract:[ Objective ] To investigate the current status of dose control for radiotherapy dosimetry in Shanghai. [ Methods ] Having set the dosimeters at the central point of a field which was 5 cm beneath the surface of a water phantom( 30cm×30cm×30cm ), irradiate the dosimeters with an absorbed dose of about 2Gy from a source at a distance of 100cm above the Water surface with an exposure field of 10 cm ×10 cm. Calculate the relative deviation between the stated dose and measured ones, and judge the quality with a deviation limit of + 5%. [ Results ] The deviations ranged from -6.8% to 8.1% among the 16 measurements, 87.5% of the accelerators passed the auditing. Even though 2 of the accelerators did not pass the quality test, their deviations were still less than 10%. [ Conclusion ] The current quality status of radiotherapy dosimetry is fairly good in Shanghai. However, further training of the technicians and auditing rules are still needed to be strengthened.
Keywords:radiotherapy dosimetry  quality control  glass dosimeter