首页 | 本学科首页   官方微博 | 高级检索  
     


Clinical significance of corrosion of cemented femoral stems in metal-on-metal hips: a retrieval study
Authors:Harry S. Hothi  Reshid Berber  Andreas C. Panagiotopoulos  Robert K. Whittaker  Camilla Rhead  John A. Skinner  Alister J. Hart
Affiliation:1.Institute of Orthopaedics and Musculoskeletal Science,University College London and the Royal National Orthopaedic Hospital,Stanmore,UK
Abstract:

Purpose

The clinical significance of corrosion of cemented femoral stems is unclear. The purpose of this retrieval study was to: (1) report on corrosion at the stem-cement interface and (2) correlate these findings with clinical data.

Methods

We analysed cemented stems (n?=?36) composed of cobalt-chromium (CoCr) and stainless steel (SS) in a series of revised metal-on-metal hips. We performed detailed inspection of each stem to assess the severity of corrosion at the stem-cement interface using a scale of 1 (low) to 5 (severe). We assessed the severity of corrosion at each stem trunnion and measured wear rates at the head taper and bearing surfaces. We used non-parametric tests to determine the significance of differences between the CoCr and SS stems in relation to: (1) pre-revision whole blood Co and Cr metal ion levels, (2) trunnion corrosion, (3) bearing surface wear and (4) taper material loss.

Results

The corrosion scores of CoCr stems were significantly greater than SS stems (p?p?p?=?0.521). There was no significant difference between the two stem types in relation to bearing wear (p?=?0.926) or taper wear (p?=?0.148).

Conclusions

Severe corrosion of cemented femoral stems is a common finding at our retrieval centre; surgeons should consider corrosion of CoCr stems as a potential source of metal ions when revising a hip.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏