The effect of blood pressure calibrations and transcranial Doppler signal loss on transfer function estimates of cerebral autoregulation
Authors:Deegan Brian M  Serrador Jorge M  Nakagawa Kazuma  Jones Edward  Sorond Farzaneh A  Olaighin Gearóid
Institution:Electrical & Electronic Engineering, NUI Galway, University Road, Galway, Ireland. b.deegan1@nuigalway.ie
Abstract:There are methodological concerns with combined use of transcranial Doppler (TCD) and Finapres to measure dynamic cerebral autoregulation. The Finapres calibration mechanism ("physiocal") causes interruptions to blood pressure recordings. Also, TCD is subject to signal loss due to probe movement. We assessed the effects of "physiocals" and TCD signal loss on transfer function estimates in recordings of 45 healthy subjects. We added artificial "physiocals" and removed sections of TCD signal from 5 min Finapres and TCD recordings. We also compared transfer function results from 5 min time series with time series as short as 1 min. Accurate transfer function estimates can be achieved in the 0.03-0.07 Hz band using beat-by-beat data with linear interpolation, while data loss is less than 10s. At frequencies between 0.07 and 0.5 Hz, transfer function estimates become unreliable with 5s of data loss every 50s. 2s data loss only affects frequency bands above 0.15Hz. Finally, accurate transfer function assessment of autoregulatory function can be achieved from time series as short as 1min, although gain and coherence tend to be overestimated at higher frequencies.