Evaluation of Autosensing as an Automatic Means of Maintaining a 2:1 Sensing Safety Margin in an Implanted Pacemaker
Authors:ANTONIO CASTRO  REAS LIEBOLD    JOSÉ VINCENTE    TIM DUNCAN  JOHN C ALLEN  Jr  AUTOSENSING INVESTIGATION TEAM
Institution:Hospital General, Jerez, Spain;Klinikum dir Universität Regensburg, Germany;Hospital Torrecardenas, Almeira, Spain;Intermedics Europe, Brussels, Belgium;Intermedics Inc., Angleton, Texas
Abstract:As intracardiac signal amplitudes fluctuate due to patient activity, drug intake, and other factors, it is imperative that an adequate sensing safety margin in an implanted pacemaker be used to compensate. We studied an investigational autosensing feature that automatically adjusts the device's sensitivity. Data were collected from 55 patients, with Intermedics model 292–03 or 294–03 pacemakers, upon inclusion of the study (Visit 1); 1 month postinclusion (Visit 2); 1 month, 1 day postinclusion (Visit 3); then 1 month, 2 days postinclusion (Visit 4). Atrial (N = 45) and/or ventricular (N = 54) thresholds were assessed at each visit; during Visit 2, myopotential tests were performed at two sensitivity settings. Autosensing was activated following Visit 1, then programmed on randomly at Visit 2 or Visit 3. From Visit 2 - Visit 4, patients were monitored during daily activities (D), exercise (E), and sleep (S) with 24-hour Holter. With Autosensing on, atrial undersensing episodes were D = 33 (p > 0.2), and S = 28 (p < 0.05); ventricular undersensing episodes were D = 6 (p > 0.5). Ventricular oversensing episodes were S = 2; atrial oversensing episodes were S = 34 (p > 0.5), D = 2, and E = 3. Comparing Autosensing adjusted sensitivity with the recommended 2:1 safety margin, 4 patients (p = 0.15) experienced atrial myopotential oversensing, and 2 patients (p = 0.15) ventricular. No unanticipated clinical events occurred. Compared with the recommended 2:1 sensing safety margin, the Autosensing feature performed equal to manual programming in preventing episodes of under/oversensing, and was better for atrial undersensing during sleep. Autosensing obviates the need for periodic reprogramming of a fixed sensitivity value.
Keywords:automatic sensing algorithm  sensing
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号