Using the acoustic interference pattern to locate the focus of a high-intensity focused ultrasound (HIFU) transducer
Authors:Wu Chih-Ching  Chen Chiung-Nien  Ho Ming-Chih  Chen Wen-Shiang  Lee Po-Huang
Institution:Department of Mechanical Engineering, National Taiwan University, Taipei, Taiwan.
Abstract:One of the main problems encountered when using conventional B-mode ultrasound (US) for targeting and monitoring purposes during ablation therapies employing high-intensity focused US (HIFU) is the appearance of strong interference in the obtained diagnostic US images. In this study, instead of avoiding the interference noise, we demonstrate how we used it to locate the focus of the HIFU transducer in both in vitro tissue-mimicking phantoms and an ex vivo tissue block. We found that when the B-mode image plane coincided with the HIFU focal plane, the interference noise was maximally converged and enhanced compared with the off-focus situations. Stronger interference noise was recorded when the angle (alpha) between the US image plane and the HIFU axis was less than or equal to 90 degrees. By intentionally creating a target (group of bubbles) at the 3.5-MHz HIFU focus (7.1 mm in length and 0.7 mm in diameter), the position of the maximal noise convergence coincided well with the target. The difference between the predicted focus and the actual one (bubbles) on x and z axes (axes perpendicular to the HIFU central axis, Fig. 1) were both about 0.9 mm. For y axis (HIFU central axis), the precision was within 1.0 mm. For tissue block ablation, the interference noise concentrated at the position of maximal heating of the HIFU-induced lesions. The proposed method can also be used to predict the position of the HIFU focus by using a low intensity output scheme before permanent changes in the target tissue were made.
Keywords:High-intensity focused ultrasound  Interference  Ablation
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号