螺旋CT低剂量扫描在颅面部外伤中的应用价值
引用本文:罗焕江. 螺旋CT低剂量扫描在颅面部外伤中的应用价值[J]. 医疗卫生装备, 2009, 30(9): 81-82
作者姓名:罗焕江
作者单位:广西医科大学,第四附属医院,广西,柳州,545005
摘    要:目的:探讨不同低剂量CT扫描在颅面部外伤中的应用价值。方法:将60例颅面部外伤患者分为4组,其中1组行常规剂量扫描,其余3组行不同低剂量CT扫描,比较常规剂量组(120kV,220mA)和低剂量组(120kV,160、80、60mA)共4组的图像质量及患者所受辐射剂量。结果:低剂量(160、80mA)CT扫描所得CT图像较好,60mA组患者所受辐射剂量较低,比常规剂量组单次扫描权重剂量指数分别降低了33%、67%和73%;低剂量(160、80mA)与常规剂量组所得图像质量比较无明显差异,60mA组与常规剂量组所得图像质量比较有显著性差异。结论:可利用低剂量(80mA)对颅面部外伤进行扫描,所得的图像对疾病的诊治无影响。

关 键 词:低剂量  颅面部  体层摄影术  X线计算机

Value of Low-dose Multi-slice CT in Application of Craniofacial Trauma
LUO Huan-jiang. Value of Low-dose Multi-slice CT in Application of Craniofacial Trauma[J]. Chinese Medical Equipment Journal, 2009, 30(9): 81-82
Authors:LUO Huan-jiang
Affiliation:LUO Huan-jiang (Fourth Hospital Affiliated to Guangxi Medical University, Liuzhou 545005, Guangxi Province, China)
Abstract:Objective To explore the different low-dose CT scanning of the eraniofacial trauma of application value.Methods 60 cases of craniofaeial trauma patients were divided into one group of conventional-dose group of line scan and 3 different low-dose CT scan, the conventional dose group(120 kV, 220 mA), low dose group (120 kV, 60, 80,60 mA) compared the level 4 group of patients with image quality and radiation dose received,Results The low-dose (160,100 mA) CT scan obtained better CT image, patients of 60 mA group received a lower radiation dose. Group single-scan dose index weights were reduced 33%, 67% and 73% respectively; There was no significant differences between low-dose group(160, 80 mA) and the normal group, while there was significant differences between the group of 60mA and normal group. Concluaion The use of low-dose (80 mA) for craniofacial trauma scan images obtained from the diagnosis and treatment of disease has no bad effect.[Chinese Medical Equipment Joumal,2009,30(9):81-82]
Keywords:low-dose  craniofacial  tomography  X-ray computer