The influence to the radiation quality by the wedge filter
HU Jie,TAO Jian min,ZHANG Ying,et al..The influence to the radiation quality by the wedge filter[J].Chinese Journal of Medical Physics,1999,16(2):67-69.
Authors:HU Jie  TAO Jian min  ZHANG Ying  
Abstract:The presence of wedge filter in the photo field will cause the change of beam quality which results in the modification of depth dose distribution comparing with that in the open field.This study iams at quantitation of these phenomena and finds out the diferences of 2.9% in beam quality index and of 4 mm in Dmax.We suggest the measurement of wedge field depth dose for the accurate dosimetry.
Keywords:wedge filter  radiation quality  dose distribution  