首页 | 本学科首页   官方微博 | 高级检索  
     


Antiinflammatory effect of heparin-coated circuits with leukocyte-depleting filters in coronary bypass surgery
Authors:Hamada Y  Kawachi K  Nakata T  Kohtani T  Takano S  Tsunooka N
Affiliation:Department of Surgery II, Ehime University School of Medicine, Shigenobu, Ehime, Japan. ryoji@m.ehime-u.ac.jp
Abstract:Cardiac surgery with cardiopulmonary bypass is associated with a systemic inflammatory response. We examined combined use of heparin coating of the cardiopulmonary bypass circuit and a leukocyte-depleting arterial line filter to reduce this response. Thirty patients were allocated randomly to equal groups with a conventional circuit and arterial line filter (C group), a heparin-coated circuit with a conventional filter (H group), or a heparin-coated circuit with a leukocyte-depleting arterial line filter (HF group). Cytokines and respiratory function were repeatedly measured perioperatively. Plasma interleukin (IL)-6 concentrations in the HF group were lower than in the C group immediately following bypass and operation, at 4 h, and 12 h (p < 0.05). Plasma IL-8 was lower in the HF group than in the C group at 4 h (p < 0.05). The respiratory index was lower immediately after bypass in the HF group than the C group (0.61 +/- 0.2 versus 1.05 +/- 0.4, p < 0.05). Heparin-coated circuits with leukocyte-depleting filters decrease inflammatory responses and improve pulmonary function during operation.
Keywords:Open heart surgery    Coronary artery bypass surgery    Cardiopulmonary bypass    Inflammatory response