鼻腔喷雾剂给药装置的性能考察
引用本文:王东兴,钱秋玉,左艳敏,李南屏,高永良. 鼻腔喷雾剂给药装置的性能考察[J]. 中国药房, 2012, 0(41): 3933-3934
作者姓名:王东兴  钱秋玉  左艳敏  李南屏  高永良
作者单位:[1]北京军区总医院263临床部药械科,北京101149 [2]军事医学科学院毒物药物研究所,北京100850
摘    要:目的:考察所选鼻腔喷雾剂给药装置的性能,建立评价该类装置性能的方法。方法:参考2010年版《中国药典》喷雾剂项下有关要求,分别以失重率和增重率为指标考察高温(60℃)和高湿(95%)条件下装置的密闭性,以及室温、40℃、60℃条件下的每喷喷量。结果:该装置在高温和高湿条件下失重率和增重率均<1%,3种条件下以60℃时的每喷喷量不准确、差异较大。结论:该装置的密闭性良好,在40℃及以下温度环境中放置时的每喷喷量准确、精密度好,符合鼻腔给药装置的要求。本文建立的方法可用于该类装置的性能考察。

关 键 词:鼻腔喷雾剂  给药装置  性能  考察

Investigation on Performance of Delivery Devices of Nasal Spray
WANG Dong-xing,QIAN Qiu-yu,ZUO Yan-min,LI Nan-pingDept. of Pharmacy,No.,Clinical Depart- ment,General Hospital of Beijing Military Region,Beijing,China GAO Yong-liang. Investigation on Performance of Delivery Devices of Nasal Spray[J]. China Pharmacy, 2012, 0(41): 3933-3934
Authors:WANG Dong-xing  QIAN Qiu-yu  ZUO Yan-min  LI Nan-pingDept. of Pharmacy  No.  Clinical Depart- ment  General Hospital of Beijing Military Region  Beijing  China GAO Yong-liang
Affiliation:WANG Dong-xing, QIAN Qiu-yu, ZUO Yan-min, LI Nan-ping(Dept. of Pharmacy, No. 263 Clinical Depart- ment, General Hospital of Beijing Military Region, Beijing 101149, China) GAO Yong-liang (Institute of Pharmacology and Toxicology, Academy of Military Medical Sciences, Beijing 100850, China)
Abstract:OBJECTIVE: To investigate the performance of delivery device of nasal spray, and to establish the method for the performance evaluation of the device. METHODS: Referring to the requirements of nasal spray stated in Chinese Pharmacopeia (2010 edition), the airtight performance of the devices was investigated under the condition of high temperature(60 ℃ ) and high humidity (95%) using proportion of weight loss and growth as index. The spray volume per spray of the selected devices was inves- tigated under room temperature, 40 ℃ and 60 ℃, respectively. RESULTS: The proportion of weight loss and growth both were smaller than 1% under high temperature and high humidity. The spray volume per spray of the selected devices was not accurate and precise under 60 ℃. CONCLUSIONS: The devices of nasal spray are in good airtight condition; the spray volume per spray is accurate and precise at 40 ℃ or below, which confirms to the requirements for the devices of nasal administration. Established method is suitable for the performance evaluation of delivery device of nasal spray.
Keywords:Nasal spray  Delivery devices  Performance  Investigation