首页 | 本学科首页   官方微博 | 高级检索  
     


Incidental detection of paradoxical air embolism with a transoesophageal Doppler probe inserted for measuring descending aortic blood flow
Authors:Draganov J  Scheeren T W L
Affiliation:Department of Anaesthesiology, University Hospital of Duesseldorf, Heinrich-Heine-University, Moorenstrasse 5, D-40225 Düsseldorf, Germany
Abstract:We present a case of paradoxical air embolism in a patient undergoingneurosurgery of the posterior fossa in the sitting position.Paradoxical air embolism was detected in the descending aortaby a transoesophageal Doppler probe, which was inserted primarilyfor non-invasive haemodynamic monitoring. The patient sufferedno serious complications of paradoxical air embolism and recoveredwell. We suggest that paradoxical air embolism can be detectedintraoperatively not only by transoesophageal echocardiography,but also with a transoesophageal Doppler probe. Br J Anaesth 2003; 90: 520–2
Keywords:complications, embolism   embolism, air   measurement techniques, transoesophageal Doppler ultrasonography   monitoring, ultrasound