首页 | 本学科首页   官方微博 | 高级检索  
     


A simple radioaerosol generator and delivery system for pulmonary ventilation studies
Authors:M. M. Ishfaq  S. K. Ghosh  A. B. Mostafa  S. R. Hesslewood  N. R. Williams  A. J. Hickey
Affiliation:(1) Department of Physics and Nuclear Medicine, Dudley Road Hospital, Dudley Road, B18 7QH Birmingham, UK;(2) Department of Pharmacy, The University of Aston in Birmingham, UK
Abstract:Details of a simple radioaerosol generator and delivery system are presented. Aerosol streams of 99mTc-DTPA solution of different distributions were produced. The most useful distribution had an activity median aerodynamic diameter (AMAD) of 0.9 mgrm with a geometric standard deviation of 1.5. This distribution also had more than 96% of aerosol particles with aerodynamic diameter <2 mgrm. The system has been used for patient lung ventilation studies. The aerosol breathing-in period to achieve a satisfactory count rate was 1.8±0.38 min. The radioaerosol images were excellent and comparable to those obtained with 81mKr gas.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏