首页 | 本学科首页   官方微博 | 高级检索  
检索        


Elimination of errors caused by first-order aliasing in velocity encoded cine-MR measurements of postoperative jets after aortic coarctation: in vitro and in vivo validation
Authors:Christine B Henk  Stephan Grampp  Jeanette Koller  Maria Schoder  Herbert Frank  Ursula Klaar  Gregor Gomischek  Gerhard H Mostbeck
Institution:1.Department of Radiology, University of Vienna, Ludwig Boltzmann Institute for Radiological Tumor Diagnosis; W?hringergürtel 18–20, 1090 Vienna, Austria,Austria;2.Department of Internal Medicine II, Division of Cardiology, University of Vienna, W?hringergürtel 18–20, 1090Vienna, Austria,Austria;3.Department of Radiology, Hospital of the Brothers of St. John, Esterhazystrasse 26, 7000 Eisenstadt, Austria,Austria
Abstract:The aim of this study was to evaluate a velocity-encoded cine-MR (VEC-MR) sequence in measuring flow velocities up to two times the velocity encoding value (VENC) in a flow phantom and to validate the method for assessing poststenotic jet velocities in postoperative patients after aortic coarctation. In vitro, a flow phantom was used (0.5T; TR/TE: 51/8 ms, flip angle=30 degrees, FOV=280 mm, 128x256 matrix VENC 40 or 80 cm/s). On binary images, maximum flow velocities (V(max)) were calculated with a region of interest (ROI, 8 pixels). With aliasing, V(max) was calculated by VENC+(V(aliasing)). In vivo, 16 postoperative patients after aortic coarctation underwent double-oblique VEC-MR imaging through the aortic arch (ECG triggering, 16 phases/RR, TR=600-800 ms, flow-encoding cranio-caudal, VENC=2 m/s). Peak systolic velocities were measured and transthoracic Doppler echocardiography (TTDE) was performed. In vitro, there were excellent correlations for MR velocity measurements with and without aliasing ( r=0.99) and for true and MR-derived flow velocities ( r=0.99). In vivo, there was good correlation between VEC-MR and TTDE-assessed V(max) values in the aorta at the former coarctation site ( r=0.90, n=16). Aliasing occurred in 13 patients. VEC-MR is a useful modality for assessing jet velocities in the follow-up of patients after aortic coarctation. Despite of aliasing, accurate velocity measurements up to two times VENC are possible using binary images.
Keywords:MR velocity measurements Aortic coarctation Aliasing Pressure gradients
本文献已被 PubMed SpringerLink 等数据库收录!