首页 | 本学科首页   官方微博 | 高级检索  
     


The use of a Leksell-BRW adapter for linac radiosurgery as an adjunct to Gamma Knife treatment
Authors:Ekstrand Kenneth E  Hinson William H  Bourland J Daniel  deGuzman Allan F  Stieber Volker W  Tatter Steven B  Ellis Thomas L
Affiliation:Department of Radiation Oncology, Wake Forest University School of Medicine, Winston-Salem, NC, USA.
Abstract:We have investigated the use of an adapter that permits the use of a Leksell coordinate frame with a linear accelerator stereotactic radiosurgery system based on the Brown-Robert-Wells (BRW) design. This device is useful when lesions that are planned for treatment on a Leksell Gamma Knife system are found to be inaccessible to the Gamma Knife. We have found that with this device objects within a head phantom can be targeted by the linear accelerator within an accuracy of approximately 1 mm.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号