PoleStar N20术中磁共振神经导航系统的导航准确性研究
引用本文:姚成军,毛颖,吴劲松,庄冬晓,周良辅. PoleStar N20术中磁共振神经导航系统的导航准确性研究[J]. 中国微侵袭神经外科杂志, 2007, 12(3): 120-124
作者姓名:姚成军  毛颖  吴劲松  庄冬晓  周良辅
作者单位:上海神经外科临床医学中心,复旦大学附属华山医院神经外科,上海,200040
基金项目:卫生部临床学科重点项目;上海市科委资助项目
摘    要:目的 研究PoleStar N20术中磁共振影像(iMRI)神经导航系统导航的准确性及影响因素,为临床应用提供指导。方法 应用塑料-有机玻璃立柱插板体模,分别测定不同空间靶点、手术体位、扫描序列、噪音水平、磁体移动等对导航准确性的影响。结果 导航准确性与扫描序列、噪音水平、扫描层厚呈显著性相关,与靶点空间位置、体位、手术床和磁体移动等不相关。结论 PoleStar N20 iMRI神经导航系统导航准确性可靠,受外界影响小,可满足临床应用所需。

关 键 词:PoleStarN20系统  术中磁共振成像  神经导航  体模,显像术
文章编号:1009-122X(2007)03-0120-05
收稿时间:2007-01-29
修稿时间:2007-02-13

Accuracy of PoleStar N-20 intraoperative MRI for neuronavigation: experimental study
YAO Chengjun, MAO Ying, WU Jinsong, et al. Accuracy of PoleStar N-20 intraoperative MRI for neuronavigation: experimental study[J]. Chinese Journal of Minimally Invasive Neurosurgery, 2007, 12(3): 120-124
Authors:YAO Chengjun   MAO Ying   WU Jinsong   et al
Abstract:Objective To investigate the accuracy of PoleStar N-20 intraoperative MRI(iMRI) for neuronavigation and its influence factors for providing theoretical guidance for clinical application.Methods A peg-board phantom was used as a model.The accuracy of neuronavigation in PoleStar N-20 iMRI was measured with reference to the different spatial target points,body position,imaging sequences,noise levels,and moving of the operating table and magnet poles.The data were analyzed statistically.Results The accuracy of PoleStar N20 iMRI was obviously associated with imaging sequence,noise level and slice thickness,but there was no correlation of the accuracy to the spatial target point,body position moving of operating table or magnet poles.Conclusion Accuracy of PoleStar N20 iMRI in neuronavigation was reliable and less influenced by outside factors,thus the iMRI system can meet clinical need.
Keywords:Polestar N20 system  intraoperative magnctic resonance imaging,neuronavigation  phantom,imaging