首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of the automated COBAS AMPLICOR hepatitis C virus PCR system.
Authors:M Poljak   K Seme     S Koren
Abstract:To evaluate the reliability and feasibility of the automated Roche COBAS AMPLICOR PCR system for routine detection of hepatitis C virus (HCV) RNA, a total of 405 serum samples previously tested by an in-house nested PCR and manual Roche AMPLICOR microwell plate HCV test were examined. Complete concordance was found between the results with the HCV COBAS AMPLICOR system and the previously determined HCV RNA status. The automated HCV COBAS AMPLICOR system provides the clinical microbiology laboratory with a specific and sensitive PCR method for rapid and reliable detection of HCV RNA.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏