不同角膜厚度下非接触眼压计与Schiotz眼压计测量眼压的比较
引用本文:安晓巨,陈开建.不同角膜厚度下非接触眼压计与Schiotz眼压计测量眼压的比较[J].国际眼科杂志,2012,12(7):1337-1339.
作者姓名:安晓巨  陈开建
作者单位:第三军医大学大坪医院野战外科研究所眼科,中国重庆市,400042
摘    要:目的:比较不同角膜厚度下非接触眼压计与Schiotz眼压计测量结果的差异。 方法:对314例627眼分别用角膜测厚仪测量角膜厚度,非接触眼压计与Schiotz眼压计测量眼压,根据角膜厚度分为:较薄组(角膜厚度≤530μm)、正常组(530μm<角膜厚度≤570μm)、较厚组(角膜厚度>570μm)。采用配对t检验比较不同角膜厚度下两种测量结果的差异,并通过Bland-Altman进行一致性评价。 结果:角膜厚度较薄组170眼,正常组301眼,较厚组156眼。两种方法在较薄组、正常组、较厚组三组中的眼压值分别为12.82±2.67,12.84±2.37mmHg;13.67±2.66,1358±2.41mmHg;15.45±2.91,14.76±2.39mmHg。两种方法测量结果在角膜厚度正常和偏薄组中差异无统计学意义,偏厚组中差异有统计学意义。 结论:在角膜厚度≤570μm的人群中,眼压的检查可在非接触眼压计和Schiotz眼压计间选择一种检查;对于角膜厚度>570μm的人群,需参考两种方法检查结果,进一步检查排除青光眼。

关 键 词:非接触眼压计  Schiotz眼压计  眼压  角膜厚度
收稿时间:2012/3/30 0:00:00
修稿时间:6/6/2012 12:00:00 AM

Comparison of the intraocular pressure of non-contact tonometry and Schiotz tonometer in different corneal thickness
Xiao-Ju An and Kai-Jian Chen.Comparison of the intraocular pressure of non-contact tonometry and Schiotz tonometer in different corneal thickness[J].International Journal of Ophthalmology,2012,12(7):1337-1339.
Authors:Xiao-Ju An and Kai-Jian Chen
Institution:Department of Ophthalmology, Research Institute of Field Surgery, Da Ping Hospital, Third Military Medical University, Chongqing 400042, China;Department of Ophthalmology, Research Institute of Field Surgery, Da Ping Hospital, Third Military Medical University, Chongqing 400042, China
Abstract:AIM:To compare the difference of intraocular pressure of non-contact tonometry and Schiotz tonometer in different corneal thickness.METHODS:Totally 314 cases(627 eyes) were recruited.Corneal thickness was measured by corneal pachymeter and intraocular pressure was measured by non-contact tonometry and Schiotz tonometer.Eyes were divided into 3 groups based on corneal thickness:thin group:corneal thickness<530μm,normal group:530μm570μm.The statistical significance of differences was evaluated by paired t-test and the agreement was assessed by Bland-Altman analysis.RESULTS:The eyes of thin,normal and thick group were:170,301 and 156 eyes,respectively.The intraocular pressure of non-contact tonometry and Schiotz tonometer in 3 groups were 12.82±2.67 mmHg,12.84±2.37mmHg;13.67±2.66mmHg,13.58±2.41mmHg;15.45±2.91mmHg,14.76±2.39mmHg,respectively.There was significant difference in thick group and no significant difference in other groups.CONCLUSION:It is suggested that either non-contact tonometry or Schiotz tonometer could measure intraocular pressure in eyes with corneal thickness≤570μm,and consult the intraocular pressure of the two measurements in eyes with corneal thickness >570μm.
Keywords:non-contact tonometry  Schiotz tonometer  intraocular pressure  corneal thickness