基于红外成像原理的睑板腺图像量化分析系统
引用本文:骆仲舟,邓宇晴,王耿媛,黄远聪,林伟锋,袁进.基于红外成像原理的睑板腺图像量化分析系统[J].眼科学报,2021(1):30-37.
作者姓名:骆仲舟  邓宇晴  王耿媛  黄远聪  林伟锋  袁进
作者单位:中山大学中山眼科中心工程中心
摘    要:目的:分析人眼的睑板腺形态学特征,探索睑板腺分析系统在眼表疾病的应用研究.方法:中山眼科中心入组正常受试者24例(42眼),进行睑板腺红外摄影.选取受试者中的10例(20眼)在同型号的设备上由二名操作员分别进行睑板腺红外摄影.图像通过自行设计的分析软件对上睑结膜中央5条腺体形态学参数进行定量分析,对数据进行重复性测试....

关 键 词:睑板腺  睑板腺形态学参数  睑板腺腺体功能评估  红外成像

Meibomian gland image quantitative analysis system based on infrared imaging principle
LUO Zhongzhou,DENG Yuqing,WANG Gengyuan,HUANG Yuancong,LIN Weifeng,YUAN Jin.Meibomian gland image quantitative analysis system based on infrared imaging principle[J].Eye Science,2021(1):30-37.
Authors:LUO Zhongzhou  DENG Yuqing  WANG Gengyuan  HUANG Yuancong  LIN Weifeng  YUAN Jin
Institution:(Engineering Center,Zhongshan Ophthalmic Center,Sun Yat-sen University,Guangzhou 510000,China)
Abstract:Objective:To analyze the morphological characteristics of meibomian glands in human eyes and to explore the application research of meibomian glands analysis system in ocular surface diseases.Methods:A total of 24 healthy subjects were recruited by Zhongshan Ophthalmic Center to infrared photography of meibomian glands.Ten of healthy subjects were selected by the two operators for infrared photography of meibomian glands on the same model of equipment.The images were repeatedly measured and analyzed by the self-designed analysis software on the morphological measurements of the five glands in the center of the upper eyelid.Results:The measured biological parameters are shown below:the average gland diameter was(0.48±0.09)mm,the average gland length was(5.25±0.68)mm,the average gland area was(2.12±0.53)mm,the gland deformation coefficient was 10.01±3.85,the development value was 6.32±1.23,the percentage of the five central glands in the central area was(10.94±2.20)%,and the glands accounted for(58.07±8.13)%of the upper conjunctiva area.There was no statistical difference between the two measurements of each index(P>0.05).Repeatability analysis results showed that coefficients of variation(CV)of all biological parameters of glands were less than 5%and the intraclass correlation coefficient(ICC)in both groups were greater than 0.95.Conclusion:The Meibomian Gland Bioimage Analyzer provides good reliability and consistency for morphological measurements of the meibomian gland,and it is expected to provide new non-invasive indicators for clinical assessment of the meibomian glands function.
Keywords:meibomian glands  meibomian gland morphological parameters  meibomian gland function evaluation  infrared imaging