首页 | 本学科首页   官方微博 | 高级检索  
     


Test-retest variability and reliability of 123I-IBZM SPECT measurement of striatal dopamine D2 receptor availability in healthy volunteers and influence of iterative reconstruction algorithms
Authors:Catafau Ana M  Bullich Santiago  Danús Mónica  Penengo María M  Cot Albert  Abanades Sergio  Farré Magí  Pavía Javier  Ros Domènec
Affiliation:Experimental Medicine, Clinical Pharmacology Discovery Medicine, Psychiatry Centre of Excellence for Drug Discovery, GlaxoSmithKline, Barcelona, Spain. ana.m.catafau@gsk.com
Abstract:OBJECTIVE: 123I-IBZM single photon emission computed tomography (SPECT) is a widely used method to measure D(2) receptor availability. However, test-retest variability and reliability have not been reported yet. This study aimed to further characterize 123I-IBZM SPECT in healthy volunteers (HVs), by assessing (1) pseudoequilibrium interval after bolus injection; (2) normal specific uptake ratio (SUR) values using filtered-backprojection (FBP); and the iterative reconstruction algorithm ordered-subsets expectation maximization (OSEM); (3) test-retest variability and reliability (intraclass correlation coefficient); and (4) influence of OSEM on test-retest variability and reliability. METHODS: Ten HVs (Group A) were scanned twice 48 h apart for test-retest variability and reliability measurements, and n = 4 of them were sequentially scanned over time. Eighteen HVs (Group B) were scanned once at pseudoequilibrium. For reconstruction FBP was used. Test-retest scans were reconstructed in addition using OSEM. SPECT-MRI coregistration was used for region of interest drawing. RESULTS: Pseudoequilibrium was achieved at 90 min postinjection (p.i.) and maintained until the end of the SPECT session (n = 4), and mean SUR at this time point was 0.96 +/- 0.14 (Groups A + B, n = 28). Mean SUR at test was 0.96 +/- 0.19 and at retest 0.94 +/- 0.19 (Group A, n = 10). Using FBP, test-retest variability was (12.7 +/- 9.6)% and reliability was 0.74. Using OSEM with 18 equivalent iterations, test-retest variability and reliability were improved to (6.5 +/- 5.2)% and 0.84, respectively. CONCLUSIONS: 123I-IBZM SPECT imaging using the bolus injection and a single scan at 90 min p.i. is a reproducible method showing acceptable test-retest variability and reliability. Test-retest variability and reliability can be substantially improved using OSEM with 12-36 equivalent iterations.
Keywords:ordered‐subsets expectation maximization  filtered‐backprojection
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号