Body-Fix 装置对脊椎转移瘤大分割调强放疗摆位误差的影响
引用本文:章雅琴,刘裕杰,许青,朱国培,章真.Body-Fix 装置对脊椎转移瘤大分割调强放疗摆位误差的影响[J].中华放射肿瘤学杂志,2013,22(4):315-317.
作者姓名:章雅琴  刘裕杰  许青  朱国培  章真
作者单位:200032 上海, 复旦大学上海医学院肿瘤学系 复旦大学附属肿瘤医院放疗科
摘    要:目的 观察Body-Fix装置用于脊椎转移瘤大分割调强放疗的误差改变,为广泛开展骨转移大分割治疗提供参考资料。方法 选择2008—2009年在本院接受大分割调强放疗的6例鼻咽癌椎体转移患者,均使用Body-Fix 装置进行固定、摆位和治疗。利用千伏级锥形束CT采集放疗前、摆位纠正后、放疗后图像,与计划CT图像进行配准,对摆位纠正后和放疗后位置误差变化情况(分次内误差)进行统计分析。结果 上、中、下段椎体左右方向分次内误差分别为(-0.6±0.5)、(-0.1±1.0)、(0.0±0.4) mm,头脚方向分次内误差分别为(1.0±1.4)、(4.8±5.7)、(0.0±0.3) mm,前后方向分次内误差分别为(1.2±5.2)、(-0.3±0.3)、(0.0±0.5) mm。结论 采用Body-Fix装置后各节段椎体的分次内误差基本能控制在2 mm内,为采用大分割调强放疗治疗椎体转移瘤提供了良好的固定保障。

关 键 词:Body-Fix装置  摆位误差  椎体转移瘤/调强放射疗法  
收稿时间:2012-07-31

The clinical effect of Body-Fix device in improving the positioning accuracy of the hypofractionated radiotherapy in the vertebral metastatic patients
ZHANG Ya-qin,LIU Yu-jie,XU Qing,ZHU Guo-pei,ZHANG Zhen.The clinical effect of Body-Fix device in improving the positioning accuracy of the hypofractionated radiotherapy in the vertebral metastatic patients[J].Chinese Journal of Radiation Oncology,2013,22(4):315-317.
Authors:ZHANG Ya-qin  LIU Yu-jie  XU Qing  ZHU Guo-pei  ZHANG Zhen
Institution:Department of Radiation Oncology, Cancer Hospital of Fudan University, Department of Oncology, Shanghai Medical College of Fudan University, Shanghai 200032,China
Abstract:Objective To study the immobilization effect and keep the accurate treating position of Body-Fix device in the patients with vertebral metastatic tumor treated by hypofractionated intensity-modulated radiotherapy. Methods From October 2008 to February 2010, six nasopharyngeal carcinoma patients with 10 treated lesion with vertebral metastasis who were treated by hypofractionated intensity-modulated radiotherapy and immobilized by the Body-Fix device were enrolled in this study. Three sets cone beam CT images were taken and recorded when patient was underway the initial setup, position correction and after radiation delivery. Comparing these images with the planning CT images to get the setup errors and the intrafractional position shifting, and the immobilization effect of Body-Fix device was analyzed. Results In the upper, middle and lower sections of the vertebrae, the intrafractional setup errors in the left-right direction were (-0.6±0.5) mm,(-0.1±1.0) mm,(0.0±0.4) mm, with in the superior-inferior direction (1.0±1.4) mm,(4.8±5.7) mm,(0.0±0.3) mm and in the anterior-posterior direction (1.2±5.2) mm,(-0.3±0.3) mm,(0.0±0.5) mm, respectively. Conclusions With Body-Fix device, the intrafractional setup errors can be minimized within 2 mm which make the accurate spinal radiosurgery technique possible.
Keywords:Body-Fix device  Set-up error  Vertebral metastasis neoplasms/intensity-modulated radiotherapy  