首页 | 本学科首页   官方微博 | 高级检索  
     


A visual probe localization and calibration system for cost-effective computer-aided 3D ultrasound
Authors:Ali Aziah  Logeswaran Rajasvaran
Affiliation:Faculty of Information Technology, Multimedia University, 63100 Cyberjaya, Malaysia.
Abstract:The 3D ultrasound systems produce much better reproductions than 2D ultrasound, but their prohibitively high cost deprives many less affluent organization this benefit. This paper proposes using the conventional 2D ultrasound equipment readily available in most hospitals, along with a single conventional digital camera, to construct 3D ultrasound images. The proposed system applies computer vision to extract position information of the ultrasound probe while the scanning takes place. The probe, calibrated in order to calculate the offset of the ultrasound scan from the position of the marker attached to it, is used to scan a number of geometrical objects. Using the proposed system, the 3D volumes of the objects were successfully reconstructed. The system was tested in clinical situations where human body parts were scanned. The results presented, and confirmed by medical staff, are very encouraging for cost-effective implementation of computer-aided 3D ultrasound using a simple setup with 2D ultrasound equipment and a conventional digital camera.
Keywords:3D ultrasound system   Camera calibration   Ultrasound probe calibration   3D visual tracking
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏