Electrocardiographic features of atrial depolarization during pacing of right atrial appendage. |
| |
Authors: | N Pandit K K Sethi J C Mohan R Arora M Khalilullah |
| |
Affiliation: | Department of Cardiology, G.B. Pant Hospital, New Delhi. |
| |
Abstract: | Identification of atrial capture during pacing from right atrial appendage is frequently difficult. Electrocardiograms of forty five patients implanted with AAI/DDD pacemakers (thirty unipolar, fifteen bipolar) were analysed to characterize the specific morphology of paced P waves. Compared to sinus P waves, atrial pacing resulted in atrial depolarization of lower amplitude (0.16 +/- 0.05 mv vs 0.11 +/- 0.032 mv, P less than 0.005) but increased duration (0.07 +/- 0.009 sec vs 0.08 +/- 0.017 sec, P less than 0.005). P wave morphology was similar in unipolar and bipolar pacing units. It was positive in lead I (80%), II (71.11%), III (80%) and aVF (75.55%). In lead aVL, paced P waves were usually diphasic with an initial negative deflection (35.55%). Precordial leads showed paced atrial depolarization of small amplitude and did not help in identification of atrial capture. In unipolar pacing P waves were best seen in lead III because of small pacing spike in this lead. Lead II was suitable for identification of paced P waves in bipolar pacing. Thus careful examination of standard ECG leads for paced P waves of low amplitude, prolonged duration and specific morphology can help in confirming atrial capture following pacing stimulus from right atrial appendage. |
| |
Keywords: | |
|
|