首页 | 本学科首页   官方微博 | 高级检索  
检索        


Preclinical evaluation of a new hollow fiber silicone membrane oxygenator for pediatric cardiopulmonary bypass: ex-vivo study.
Authors:Shinji Kawahito  Shuji Haraguchi  Tomohiro Maeda  Tadashi Motomura  Tamaki Takano  Kenji Nonaka  Joerg Linneweber  Seiji Ichikawa  Masaki Kawamura  Hiroshi Ishitoya  Julie Glueck  Koshiro Sato  Yukihiko Nosé
Institution:Michael E. DeBakey Department of Surgery, Baylor College of Medicine, Houston, USA.
Abstract:Based on the results of many experimental models, a hollow fiber silicone membrane oxygenator applicable for long-term extracorporeal membrane oxygenation (ECMO) was developed. For further high performance and antithrombogenicity, this preclinical model was modified, and a new improved oxygenator was successfully developed. In addition to ECMO application, the superior biocompatibility of silicone must be advantageous for pediatric cardiopulmonary bypass (CPB). An ex vivo short-term durability test for pediatric CPB was performed using a healthy miniature calf for six hours. Venous blood was drained from the left jugular vein of a calf, passed through the oxygenator and infused into the left carotid artery using a Gyro C1E3 centrifugal pump. For six hours, the O2 and CO2 gas transfer rates were maintained around 90 and 80 ml/min at a blood flow rate of 2 L/min and V/Q=3, respectively. The plasma free hemoglobin was maintained around 5 mg/dl. These data suggest that this newly improved oxygenator has superior efficiency, less blood trauma, and may be suitable for not only long-term ECMO but also pediatric CPB usage.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏