Hazards of non coherent light sources as determined by the framework of IEC TR-60825-9 |
| |
Authors: | DM Clarkson |
| |
Affiliation: | 1. Department of Clinical Physics and Bioengineering, Radiotherapy Centre, Walsgrave Hospital, UHCW NHS Trust, Coventry, CV2 2DX, UKdouglas.clarkson@uhcw.nhs.uk |
| |
Abstract: | An overview is provided of the scope and application of IEC TR 60825-9 in determination of hazards of non coherent optical radiation sources. Specific areas reviewed in detail include those relating to hazards of ultraviolet radiation, retinal thermal hazard and blue light photochemical hazard. The effect of spectral temperature on relative risk relating to retinal thermal hazard and blue light photochemical hazard are outlined. Assessment of risk factors of these hazards was determined using a range of measurement devices including spectral radiometer, broad band power meter and a locally developed device for determination of pulse profile of pulsed light sources. |
| |
Keywords: | incubator cot neonates |
|
|