首页 | 本学科首页   官方微博 | 高级检索  
检索        


Microwave sensor for the investigation of glucose-dependent reflection properties in aqueous samples
Authors:Satish Singh  Rajesh Singh  Kushal Sen  Sneh Anand
Institution:1. Centre for Biomedical Engineering, Indian Institute of Technology, New Delhi, India;2. LF, HF Impedance and DC Metrology, CSIR-National Physical Laboratory, New Delhi, India;3. singhsp3@nplindia.org;5. Bharti School of Telecom Technology and Management, Indian Institute of Technology, New Delhi, India;6. Department of Textile Technology, Indian Institute of Technology, New Delhi, India
Abstract:Abstract

The reported paper presents the development of a microwave sensor with a resonant frequency 2.4?GHz. The sensor is also demonstrated in vitro to investigate the variation in its response as a function of glucose concentration. The sensor could be used for the monitoring of blood glucose level in diabetics through non-invasive technology. The approach followed is based on the notion that, change in glucose concentration in the blood affects dielectric properties of blood which in turn produce an impact on reflection properties of the sensor. This effect on response of sensor will be ultimately used to estimate blood glucose concentration. The design specifications considered for the development of sensor are defined in the paper. The experimental setup for in vitro experiment and procedure employed for the investigation of the reflection properties of the sensor as a function of glucose concentration are also discussed in detail. The shift in resonance frequency and the change in the magnitude of the reflection coefficient of proposed sensor have been observed. The reported measurement results are the preliminary results in exploring the implementation of proposed sensor for non-invasive blood glucose monitoring.
Keywords:Glucose concentration  microwave sensor  non-invasive  reflection coefficient  resonant frequency
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号