Beat to beat analysis of mechanical heart valves by means of return map
Authors:M Grigioni  C Daniele  C Del Gaudio  U Morbiducci  G D'avenio  D Di Meo
Institution:1. Technology and Health Department, Istituto Superiore di Sanità, Viale Regina Elena, 299, 00161, Rome, Italygrigioni@iss.it;3. Technology and Health Department, Istituto Superiore di Sanità, Viale Regina Elena, 299, 00161, Rome, Italy
Abstract:Three mechanical heart valves (two bileaflet prostheses and a tilting one) were investigated in a basic hardware setup in order to evaluate with a hydrophone their opening and closing action in time and in amplitude of each beat. The recorded signal was then segmented into the series of cycles xi(t) having a temporal duration equal to the working period imposed on the valve. Two return maps were defined, in order to evaluate the degree of dispersion of the resulting scatter plot: (i) the amplitude map xi(t) versus xi+1(t); (ii) the delay map for the closure of the valve within each beat versus the successive ones. To evaluate the results obtained, two indices were proposed based on both the degree of dispersion and the deviation of the regression line of the resulting scatter plot with respect to the bisector of the map plane. The tilting disc valve showed a lower degree of dispersion, both in the amplitude signal and in the closure time delays, with respect to the other two bileaflet heart valves. The methodology proposed here could be regarded as an alternative non-invasive tool to investigate the dynamic behaviour of prosthetic heart valves, especially in the case of their suspected failure.
Keywords: