Development of a human DNA quantitation system |
| |
Authors: | Mandrekar M N Erickson A M Kopp K Krenke B E Mandrekar P V Nelson R Peterson K Shultz J Tereba A Westphal N |
| |
Affiliation: | Research and Development, Genetic Analysis, Promega Corporation, 2800 Woods Hollow Road, Madison, WI 53711-5399, USA. mmandrek@promega.com |
| |
Abstract: | The AluQuant Human DNA Quantitation System has been developed for human-specific quantitation of forensic samples. This system uses probes specific to repetitive genetic elements allowing quantitation without target amplification. Target immobilization is unnecessary with employment of solution hybridization. The AluQuant Human DNA Quantitation System uses a series of enzymatic reactions to produce a luminescent signal proportional to the quantity of human DNA present. This report demonstrates a range of quantitation from 0.1-50 ng of human DNA. Signal from non-human DNAs tested was insignificant and addition of non-human DNAs into a human sample did not alter quantitation. Lastly, the system was unaffected by degradation of sample through sonication. The AluQuant Human DNA Quantitation System is a simple and sensitive method for quantitating the concentration of human DNA in forensic samples. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|