首页 | 本学科首页   官方微博 | 高级检索  
     


A randomized, controlled, prospective study validating the acquisition of percutaneous renal collecting system access skills using a computer based hybrid virtual reality surgical simulator: phase I
Authors:Knudsen Bodo E  Matsumoto Edward D  Chew Ben H  Johnson Brooke  Margulis Vitaly  Cadeddu Jeffrey A  Pearle Margaret S  Pautler Stephen E  Denstedt John D
Affiliation:Division of Urology, Ohio State University Medical Center, Columbus, Ohio 43210, USA. bodo.knudsen@osumc.edu
Abstract:PURPOSE: The need to develop new methods of surgical training combined with advances in computing has led to the development of sophisticated virtual reality surgical simulators. The PERC Mentortrade mark is designed to train the user in percutaneous renal collecting system access puncture. We evaluated and established face, content and construct validation of the simulator in this task. MATERIALS AND METHODS: A total of 63 trainees underwent baseline testing on the simulator, consisting of percutaneous renal puncture followed by the introduction of a guidewire into the collecting system. Subjects were then randomized to an intervention arm, in which they underwent 2, 30-minute training sessions on the simulator, and a control arm, in which no further training was given, followed by repeat testing. Performance was assessed using a global rating scale and by virtual reality derived parameters. RESULTS: There were no significant differences between the 2 groups with respect to baseline measures. Subjects who underwent training with the simulator demonstrated significant improvement in objective and subjective parameters compared to their baseline performance and compared to the untrained control group. Spearman rank correlations demonstrated a significant relationship between multiple parameters of the objective and subjective data. CONCLUSIONS: Training on the simulator improves virtual reality skills. It may allow trainees to develop the basic skills necessary to perform percutaneous renal collecting system access. Face and content validity were demonstrated and construct validity was supported by establishing convergent validity.
Keywords:kidney   nephrostomy   percutaneous   computer simulation   professional competence
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号