中央角膜厚度对3种眼压计眼压测量值的影响
引用本文:周和政,张绍阳. 中央角膜厚度对3种眼压计眼压测量值的影响[J]. 眼科新进展, 2011, 0(7): 681-683
作者姓名:周和政  张绍阳
作者单位:广州军区武汉总医院眼科;广西壮族自治区人民医院眼科;
摘    要:目的比较分析不同中央角膜厚度(central corneal thickness,CCT)下Icare回弹式眼压计、Goldmann压平式眼压计(Goldmann applanation tonometer,GAT)和动态轮廓眼压计(dynamic contour tonometry,DCT)的眼压测量结果,探讨CCT对3种眼压计测量值的影响。方法 对78例患者152眼分别用Icare、GAT、DCT3种眼压计进行眼压测量,并进行CCT的测量,对比不同CCT下3种眼压计的测量结果,分析眼压测量值与CCT的关系。结果 在全部受测者中Icare、GAT、DCT测得的眼压均值分别为(19.16±5.03)mmHg(1 kPa=7.5 mmHg)、(18.41±4.52)mmHg和(17.23±3.69)mmHg,三者之间有显著差异(F=7.256,P=0.001)。Icare和GAT的眼压测量值均与CCT显著相关(r=0.341,P<0.001;r=0.333,P<0.001),CCT每改变10μm,Icare的眼压值改变0.47 mmHg,GAT的眼压值改变0.41 mmHg;而DCT的眼压测量值与CCT无显著相关(r=0.032,P=0.699)。结论 Icare、GAT的眼压测量值均明显受CCT的影响,而Icare受CCT影响的程度较GAT的稍大,DCT的眼压测量值基本不受CCT的影响。

关 键 词:Icare回弹式眼压计  Goldmann压平式眼压计  动态轮廓眼压计  眼压  中央角膜厚度

Influence of central corneal thickness on intraocular pressure measured with three tonometers
ZHOU He-Zheng,ZHANG Shao-Yang From the. Influence of central corneal thickness on intraocular pressure measured with three tonometers[J]. Recent Advances in Ophthalmology, 2011, 0(7): 681-683
Authors:ZHOU He-Zheng  ZHANG Shao-Yang From the
Affiliation:ZHOU He-Zheng,ZHANG Shao-Yang From the Department of Ophthalmology,Wuhan General Hospital of Chinese PLA,Guangzhou Military Command,Wuhan 430070,Hubei Province,China,Department of Ophthalmology,the People's Hospital of Guangxi Zhuang Autonomous Region,Nanning 530021,Guangxi Zhuang Autonomous Region
Abstract:Objective To evaluate the influence of central corneal thickness(CCT)on intraocular pressure(IOP)values measured with Icare rebound tonometer(Icare),Goldmann applanation tonometer(GAT)and PASCAL dynamic contour tonometer(DCT).Methods CCT and IOP were measured in 152 eyes of 78 cases.IOP measurement was performed on each eye with Icare,GAT and DCT.The IOP values with three tonometers under different CCT were compared,and the relations between IOP and CCT were analyzed.Results The mean IOP values measured wit...
Keywords:Icare rebound tonometer  Goldmann applanation tonometer  dynamic contour tonometer  intraocular pressure  central corneal thickness  